Centrifuge safety: balancing, rotor care and logbooks

Nofira editorial team 4 min read

  • centrifuge
  • safety
  • maintenance

Centrifuge safety rests on four habits: balance opposite tubes by mass, inspect the rotor before every use, stay inside the rotor's speed and lifetime limits, and keep a logbook. A failed rotor releases a great deal of energy, so these checks are not optional.

How do you balance a centrifuge correctly?

Place tubes opposite each other, filled to the same level, with equal mass within the tolerance in the manual. Balance by weighing, not by eye, because liquids of different density with equal volume differ in mass. With an odd number of samples, add a balance tube filled with water or a suitable liquid. In swing-bucket rotors, load all buckets, and keep the buckets matched with their caps and positions.

What should you check before each run?

  1. Tubes and adapters are rated for the speed, the relative centrifugal force and the sample.
  2. Rotor and buckets are clean and dry, with no visible damage.
  3. The rotor is seated properly and its lid or locking nut is closed as the manual says.
  4. Speed and time are set within the rotor's rated limit.
  5. The lid is closed.

Stay with the instrument until it reaches speed, and stop it at once if you hear or feel unusual noise or vibration. Wait until the rotor has stopped before you open the lid.

How do you care for a rotor?

Clean it after use with the agents the manufacturer permits, then dry it, because residues and moisture start corrosion. Avoid aggressive chemicals. Inspect it regularly for corrosion, cracks, scratches and pitting, especially in the tube cavities, and take it out of service if you find any. Do not use a rotor on a different centrifuge unless the manufacturer says it fits.

What about infectious or hazardous samples?

Use sealed tubes and aerosol-tight rotor lids or buckets and open them in a suitable containment, such as a biosafety cabinet. Which measures are required depends on the risk assessment, so follow your safety officer. See also biosafety cabinet or fume hood.

Why keep a rotor logbook?

Many rotors have a lifetime limit that depends on speed and number of runs, and the manufacturer says how to count. A logbook per rotor records runs or hours, the cleaning, the inspections and any incidents, so you know when the limit is near and who used it. Add the centrifuge to your maintenance plan as well.

Nofira can hold a maintenance plan with reminders and a log for each device, including the centrifuge. See maintenance.

Frequently asked questions

How do you balance a centrifuge?

Place tubes opposite each other in the rotor, filled to the same level and weighing the same within the tolerance in the manual. Balance by weighing, not by eye, and use a balance tube with water for an odd number of samples.

When should a centrifuge rotor be retired?

When it shows corrosion, cracks, deep scratches or deformation, or when it reaches its lifetime limit. Lifetime limits come from the manufacturer and depend on rotor type, speed and number of runs, so check the manual.