Biosafety cabinet or fume hood: which one for which job
Use a biosafety cabinet when you work with biological agents and need to protect the sample, the person and the environment. Use a fume hood when you work with hazardous chemical vapours, gases or dusts, and the main risk is breathing them in.
What does a biosafety cabinet do?
A microbiological safety cabinet, standardised in Europe as EN 12469, draws air in at the front and passes the outflowing and the downflowing air through HEPA filters. A Class II cabinet therefore protects the operator, the sample and the environment. It is the usual choice for cell culture and work with infectious material. Which containment level and cabinet class apply depends on the risk assessment and national law.
What does a fume hood do?
A fume hood, standardised in Europe as EN 14175, pulls air from the room across the work area and exhausts it, with the sash as a barrier. It protects the person from vapours, gases and dusts, and it does not protect the sample. Use it for volatile solvents, acids, and reactions that give off fumes.
What is the main difference?
- A biosafety cabinet filters air, and filters catch particles and microorganisms, not chemical vapours.
- A fume hood moves contaminated air out of the room and does nothing for sterility.
- A biosafety cabinet keeps the work sterile, a fume hood does not.
What about a clean bench?
A laminar flow clean bench keeps the sample clean by blowing filtered air across it, and in many designs the air flows towards you. It protects the sample only, not you. Do not use a clean bench for infectious, toxic or allergenic material.
What if the work has both risks?
Work with both biological and chemical hazards needs a decision from your safety officer, who looks at the agents, the quantities and the cabinet types available. Some cabinets are ducted to the outside, which changes what they can handle, but do not decide this yourself.
How do you use either safely?
- Do not block the air intake or rear grille, and keep the work area uncluttered.
- Work well inside the cabinet or hood, away from the front edge.
- Keep movement slow and the sash at the marked height.
- Have the device tested at the intervals your rules require, and keep the certificate with the device record, see maintenance planning.
Nofira can hold the testing dates and certificates for each device and remind you before they lapse, but the choice of containment stays with your safety officer. See lab software.
Frequently asked questions
Can I use a biosafety cabinet for chemicals?
Only in very small amounts and only if the cabinet type and your safety officer allow it. Most biosafety cabinets recirculate filtered air, and HEPA filters do not remove vapours, so volatile or toxic chemicals belong in a fume hood.
Which standards apply?
EN 12469 covers microbiological safety cabinets, and EN 14175 covers fume cupboards. National rules and your institution's requirements may add to these.