Pipette calibration: how often and how to keep track
How often to calibrate pipettes depends on how much they are used, how tight your method tolerance is and what your quality system requires. Many labs set a fixed interval and shorten it for heavily used pipettes or when checks show drift.
How often should pipettes be calibrated?
There is no universal interval. A pipette used daily for a sensitive assay needs more attention than one used twice a month for buffers. Start from your quality system or accreditation, if you have one, and from the manufacturer's recommendation. If a check shows the pipette outside tolerance, shorten the interval for that pipette and look at the ones used the same way.
What is the difference between calibration and a routine check?
Calibration tests the pipette against defined limits, usually by a gravimetric method: weighing dispensed water on an analytical balance and converting the mass to volume. ISO 8655 is the standard for piston-operated volumetric apparatus and its test methods. Calibration is done by trained staff or by a service provider, and it produces a record.
A routine check is quicker and informal: check that the tip seats properly, fill a tip and hold it to see whether a drop forms, which points to a leak, and note a piston that feels rough or sticks. A routine check does not replace calibration, but it catches obvious faults between due dates.
What should a pipette record contain?
- A unique ID, engraved or labelled on the pipette, plus model and volume range.
- Date of the last calibration and who did it.
- The result and whether it passed.
- The next due date.
- Where the pipette lives, and who is responsible.
Put a small label with the ID and the next due date on the pipette itself, so anyone can see at a glance if it is overdue.
What do you do with an overdue or failed pipette?
Take it out of use and mark it clearly, for example with a tag, so nobody picks it up by mistake. Send it for service or repair, then recalibrate before it returns. If results were produced with a pipette that later failed, ask your quality lead how far back to review. That decision depends on your method and your quality rules.
How can you track due dates?
Keep one record per pipette, not one list per person, and review the due dates on a fixed day each month. A shared spreadsheet is enough for a small lab. Software that rolls the next date forward and reminds you helps once there are dozens of devices. Nofira offers recurring plans with reminders for calibration and maintenance, see maintenance and calibration. For the wider routine, read a preventive maintenance plan for lab equipment.
Frequently asked questions
How often should pipettes be calibrated?
Many labs use a fixed interval, often once or twice a year, but the right one depends on how heavily the pipette is used, how tight your method tolerance is and what your quality system or accreditation requires. Shorten the interval if checks show drift.
Which standard applies to pipette calibration?
ISO 8655 describes piston-operated volumetric apparatus and the test methods for them. Your institution may also apply its own procedure or a service provider's certificate.